!!Projects per year
Crynodeb
A power MOSFET-based push–pull configuration nanosecond-pulse generator has been designed, constructed, and characterized to permeabilize cells for biological and medical applications. The generator can deliver pulses with durations ranging from 80 ns up to 1 μs and pulse amplitudes up to 1.4 kV. The unit has been tested for in vitro experiments on a medulloblastoma cell line. Following the exposure of cells to 100, 200, and 300 ns electric field pulses, permeabilization tests were carried out, and viability tests were conducted to ver- ify the performance of the generator. The maximum temperature rise of the biological load was also calculated based on Joule heating energy conservation and experimental validation. Our results indicate that the developed device has good capabilities to achieve well-controlled electro-manipulation in vitro.
| Iaith wreiddiol | Saesneg |
|---|---|
| Tudalennau (o-i) | 645-657 |
| Cyfnodolyn | International Journal of Microwave and Wireless Technologies |
| Cyfrol | 11 |
| Rhif cyhoeddi | 7 |
| Dyddiad ar-lein cynnar | 27 Mai 2019 |
| Dynodwyr Gwrthrych Digidol (DOIs) | |
| Statws | Cyhoeddwyd - 30 Medi 2019 |
NDC y CU
Mae’r allbwn hwn yn cyfrannu at y Nod(au) Datblygu Cynaliadwy canlynol
-
NDC 7 Ynni Fforddiadwy a Glân
Ôl bys
Gweld gwybodaeth am bynciau ymchwil 'Push–pull configuration of high-power MOSFETs for generation of nanosecond pulses for electropermeabilization of cells'. Gyda’i gilydd, maen nhw’n ffurfio ôl bys unigryw.Prosiectau
- 1 Wrthi'n gweithredu
-
Semiconductor-based Ultrawideband Micromanipulation of CAncer STEm Cells - SUMCASTEC (Grant No. 737164)
Palego, C. (PY)
1/01/17 → 31/12/99
Project: Ymchwil
Traethodau Ymchwil Myfyriwr
-
High Voltage Nanosecond Pulsed Electric Field (nsPEF) Electroporation Systems
Davies, I. (Awdur), Hancock, C. (Goruchwylydd) & Palego, C. (Goruchwylydd), 20 Maw 2023Traethawd ymchwil myfyriwr: Doethur mewn Athroniaeth