Neidio i’r brif dudalen lywio Neidio i chwilio Neidio i’r prif gynnwys

Simulated performance of the in-beam conversion-electron spectrometer, SPICE

  • S. Ketelhut
  • , L. J. Evitts
  • , A. B. Garnsworthy
  • , C. Bolton
  • , G. C. Ball
  • , R. Churchman
  • , R. Dunlop
  • , G. Hackman
  • , R. Henderson
  • , M. Moukaddam
  • , E. T. Rand
  • , C. E. Svensson
  • , J. Witmer
  • TRIUMF
  • University of Guelph, Ontario

Allbwn ymchwil: Cyfraniad at gyfnodolynErthygladolygiad gan gymheiriaid

Crynodeb

The SPICE spectrometer is a new in-beam electron spectrometer designed to operate in conjunction with the TIGRESS HPGe Clover array at TRIUMF-ISAC. The spectrometer consists of a large area, annular, segmented lithium-drifted silicon electron detector shielded from the target by a photon shield. A permanent magnetic lens directs electrons around the photon shield to the detector. Experiments will be performed utilising Coulomb excitation, inelastic-scattering, transfer and fusion–evaporation reactions using stable and radioactive ion beams with suitable heavy-ion detection. Good detection efficiency can be achieved in a large energy range up to 3500 keV electron energy using several magnetic lens designs which are quickly interchangeable. COMSOL and Geant4 simulations have been used to maximise the detection efficiency. In addition, the simulations have guided the design of components to minimise the contributions from various sources of backgrounds.
Iaith wreiddiolSaesneg
Tudalennau (o-i)154-163
CyfnodolynNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
Cyfrol753
Dyddiad ar-lein cynnar29 Maw 2014
Dynodwyr Gwrthrych Digidol (DOIs)
StatwsCyhoeddwyd - 21 Gorff 2014
Cyhoeddwyd yn allanolIe

Ôl bys

Gweld gwybodaeth am bynciau ymchwil 'Simulated performance of the in-beam conversion-electron spectrometer, SPICE'. Gyda’i gilydd, maen nhw’n ffurfio ôl bys unigryw.

Dyfynnu hyn