Abstract
This work presents a rapid, calibration-free methodology for constructing high-fidelity digital twins of laboratory X-ray computed-tomography (CT) systems using only radiographs acquired with the instrument itself. By measuring the non-linear relationship between operator-set cathode current and actual anode current, we correct the emitted photon flux and derive a representative flux model through inverse optimisation of image noise. The approach incorporates filtration, scintillator response, detector point-spread function, electronic gain, quantisation, and over-exposure limits, enabling realistic photonic-noise synthesis without specialised dosimetry equipment. Validation on a dual-tube
high-energy (DTHE) micro-focus CT platform demonstrates sub-5% error in signal-to-noise ratio across a broad range of voltages (60–260 kV), currents (50–500 μA), exposures, and gains, and accurate reproduction of material-specific attenuation for aluminium, copper, and steel specimens. Comparisons with
Monte-Carlo (GATE) and analytical (gVirtualXray) simulations confirm that the inverse-derived twins achieve comparable visual and quantitative fidelity while requiring orders of magnitude less computation time (≈0.03 s per projection). Limitations stem from unknown internal filtration and gain-dependent
noise scalingNonetheless, the proposed pipeline provides a practical, reproducible framework for generating digital twins that support rapid prototyping, experimental planning, and advanced data-driven
analyses in X-ray imaging research.
high-energy (DTHE) micro-focus CT platform demonstrates sub-5% error in signal-to-noise ratio across a broad range of voltages (60–260 kV), currents (50–500 μA), exposures, and gains, and accurate reproduction of material-specific attenuation for aluminium, copper, and steel specimens. Comparisons with
Monte-Carlo (GATE) and analytical (gVirtualXray) simulations confirm that the inverse-derived twins achieve comparable visual and quantitative fidelity while requiring orders of magnitude less computation time (≈0.03 s per projection). Limitations stem from unknown internal filtration and gain-dependent
noise scalingNonetheless, the proposed pipeline provides a practical, reproducible framework for generating digital twins that support rapid prototyping, experimental planning, and advanced data-driven
analyses in X-ray imaging research.
| Original language | English |
|---|---|
| Article number | 172109 |
| Journal | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment |
| Volume | 1094 |
| Early online date | 6 Oct 2026 |
| DOIs | |
| Publication status | E-pub ahead of print - 6 Oct 2026 |
Keywords
- Digital twin
- Digital Twinning
- X-ray simulation
- Computed Tomography
- gVirtualXray
Fingerprint
Dive into the research topics of 'Creating functional digital twins of X-ray CT systems without hassle'. Together they form a unique fingerprint.Prizes
-
CoSeC Impact Award
Mitchell, I. (Recipient) & Vidal, F. (Recipient), 26 May 2023
Prize: Prize (including medals and awards)
-
Runner-up award for best student presentation
Mitchell, I. (Recipient) & Vidal, F. (Recipient), 12 Oct 2023
Prize: Prize (including medals and awards)
Impacts
-
gVirtualXRay: an open source library from Bangor University to simulate X-ray images on GPU in real-time
Vidal, F. (Participant)
Impact: Technological, Societal
Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver