Errors in H-MRS estimates of brain metabolite concentrations caused by failing to take into account tissue-specific signal relaxation
Allbwn ymchwil: Cyfraniad at gyfnodolyn › Erthygl › adolygiad gan gymheiriaid
StandardStandard
Yn: NMR in biomedicine, Cyfrol 31, Rhif 6, 06.2018, t. e3914.
Allbwn ymchwil: Cyfraniad at gyfnodolyn › Erthygl › adolygiad gan gymheiriaid
HarvardHarvard
APA
CBE
MLA
VancouverVancouver
Author
RIS
TY - JOUR
T1 - Errors in H-MRS estimates of brain metabolite concentrations caused by failing to take into account tissue-specific signal relaxation
AU - Gasparovic, Charles
AU - Chen, Hongji
AU - Mullins, Paul G
N1 - Copyright © 2018 John Wiley & Sons, Ltd.
PY - 2018/6
Y1 - 2018/6
N2 - Accurate measurement of brain metabolite concentrations with proton magnetic resonance spectroscopy (1 H-MRS) can be problematic because of large voxels with mixed tissue composition, requiring adjustment for differing relaxation rates in each tissue if absolute concentration estimates are desired. Adjusting for tissue-specific metabolite signal relaxation, however, also requires a knowledge of the relative concentrations of the metabolite in gray (GM) and white (WM) matter, which are not known a priori. Expressions for the estimation of the molality and molarity of brain metabolites with 1 H-MRS are extended to account for tissue-specific relaxation of the metabolite signals and examined under different assumptions with simulated and real data. Although the modified equations have two unknowns, and hence are unsolvable explicitly, they are nonetheless useful for the estimation of the effect of tissue-specific metabolite relaxation rates on concentration estimates under a range of assumptions and experimental parameters using simulated and real data. In simulated data using reported GM and WM T1 and T2 times for N-acetylaspartate (NAA) at 3 T and a hypothetical GM/WM NAA ratio, errors of 6.5-7.8% in concentrations resulted when TR = 1.5 s and TE = 0.144 s, but were reduced to less than 0.5% when TR = 6 s and TE = 0.006 s. In real data obtained at TR/TE = 1.5 s/0.04 s, the difference in the results (4%) was similar to that obtained with simulated data when assuming tissue-specific relaxation times rather than GM-WM-averaged times. Using the expressions introduced in this article, these results can be extrapolated to any metabolite or set of assumptions regarding tissue-specific relaxation. Furthermore, although serving to bound the problem, this work underscores the challenge of correcting for relaxation effects, given that relaxation times are generally not known and impractical to measure in most studies. To minimize such effects, the data should be acquired with pulse sequence parameters that minimize the effect of signal relaxation.
AB - Accurate measurement of brain metabolite concentrations with proton magnetic resonance spectroscopy (1 H-MRS) can be problematic because of large voxels with mixed tissue composition, requiring adjustment for differing relaxation rates in each tissue if absolute concentration estimates are desired. Adjusting for tissue-specific metabolite signal relaxation, however, also requires a knowledge of the relative concentrations of the metabolite in gray (GM) and white (WM) matter, which are not known a priori. Expressions for the estimation of the molality and molarity of brain metabolites with 1 H-MRS are extended to account for tissue-specific relaxation of the metabolite signals and examined under different assumptions with simulated and real data. Although the modified equations have two unknowns, and hence are unsolvable explicitly, they are nonetheless useful for the estimation of the effect of tissue-specific metabolite relaxation rates on concentration estimates under a range of assumptions and experimental parameters using simulated and real data. In simulated data using reported GM and WM T1 and T2 times for N-acetylaspartate (NAA) at 3 T and a hypothetical GM/WM NAA ratio, errors of 6.5-7.8% in concentrations resulted when TR = 1.5 s and TE = 0.144 s, but were reduced to less than 0.5% when TR = 6 s and TE = 0.006 s. In real data obtained at TR/TE = 1.5 s/0.04 s, the difference in the results (4%) was similar to that obtained with simulated data when assuming tissue-specific relaxation times rather than GM-WM-averaged times. Using the expressions introduced in this article, these results can be extrapolated to any metabolite or set of assumptions regarding tissue-specific relaxation. Furthermore, although serving to bound the problem, this work underscores the challenge of correcting for relaxation effects, given that relaxation times are generally not known and impractical to measure in most studies. To minimize such effects, the data should be acquired with pulse sequence parameters that minimize the effect of signal relaxation.
KW - Brain
KW - Magnetic Resonance Spectroscopy
KW - Metabolite Relaxation
U2 - 10.1002/nbm.3914
DO - 10.1002/nbm.3914
M3 - Article
C2 - 29727496
VL - 31
SP - e3914
JO - NMR in biomedicine
JF - NMR in biomedicine
SN - 0952-3480
IS - 6
ER -