Neonatal digital chest radiography- should we be using additional copper filtration?

Research output: Contribution to journalArticlepeer-review

Electronic versions

Documents

DOI

  • Jenna Ruth Tugwell-Allsup
    Betsi Cadwaladr University Health Board NHS Wales
  • Rhys Wyn Morris
    School of Healthcare Sciences, Bangor University
  • Kate Thomas
    Betsi Cadwaladr University Health Board NHS Wales
  • Richard Hibbs
    Integral Business Support Ltd
  • Andrew England
    Keele University
Objectives: Copper filtration removes lower energy X-ray photons, which do not enhance image quality but would otherwise contribute to patient radiation dose. This study explores the use of additional copper filtration for neonatal mobile chest imaging. Methods: A controlled factorial-designed experiment was used to determine the effect of independent variables on image quality and radiation dose. These variables included: copper filtration (0 Cu, 0.1 Cu and 0.2 Cu), exposure factors, source-to-image distance and image receptor position (direct / tray). Image quality was evaluated using absolute visual grading analysis (VGA) and contrast-to-noise ratio (CNR) and entrance surface dose (ESD) was derived using an ionising chamber within the central X-ray beam. Results: VGA, CNR and ESD significantly reduced (p < 0.01) when using added copper filtration. For 0.1 Cu, the percentage reduction was much greater for ESD (60%) than for VGA (14%) and CNR (20%), respectively. When compared to the optimal combinations of parameters for incubator imaging using no copper filtration, an increase in kV and mAs when using 0.1-mm Cu resulted in better image quality at the same radiation dose (direct) or, equal image quality at reduced dose (in-tray). The use of 0.1-mm Cu for neonatal chest imaging with a corresponding increase in kV and mAs is therefore recommended. Conclusion: Using additional copper filtration significantly reduces radiation dose (at increased mAs) without a detrimental effect on image quality. Advances in knowledge: This is the first study, using an anthropomorphic phantom, to explore the use of additional Cu for digital radiography neonatal chest imaging and therefore helps inform practice to standardise and optimise this imaging examination.
Original languageEnglish
Article number20211026
JournalBritish Journal of Radiology
Volume95
Issue number1130
Early online date14 Dec 2021
DOIs
Publication statusPublished - 1 Feb 2022
Externally publishedYes

Total downloads

No data available
View graph of relations